Powder Whiteness Tester
Categories:Other Lab Tester
Name: Powder Whiteness Tester
Model:PN-PWT
PN-48B Powder Whiteness Tester is designed to determine powder whiteness of inorganic chemicas products for the quality control (QC) and research and development (R&D). Kaolincalcium carbonate talc barite activated clay titanium dioxide barium s
Download:
Video:Click Play
Introduction
Technical parameter
Power supply: AC100~240V 2A 50/60Hz
Accuracy:△R457≤0.5、△Y≤1.0
Resolution:0.1
Measuring range:(1 ~ 160)%
Integrating sphere: Diffused illumination in 30mm window in 150mm diameter
Sample size:The test plate diameter ≥30mm Sample thickness≤ 40 mm
Display:128 × 64 dot matrix LCD
Working condition:Indoor temp.(20 ± 10 )°C;relative humidity < 85%
Dimension:(370 × 260 × 430)mm(L×W×H)
Net Weight: About 11.5 kg
Accuracy:△R457≤0.5、△Y≤1.0
Resolution:0.1
Measuring range:(1 ~ 160)%
Integrating sphere: Diffused illumination in 30mm window in 150mm diameter
Sample size:The test plate diameter ≥30mm Sample thickness≤ 40 mm
Display:128 × 64 dot matrix LCD
Working condition:Indoor temp.(20 ± 10 )°C;relative humidity < 85%
Dimension:(370 × 260 × 430)mm(L×W×H)
Net Weight: About 11.5 kg
Product features
To measure the Blue Reflectance Factor (R457) of samples. Blue Reflectance Factor is so called “Blue Whiteness” or “ISO Brightness”.
Adopt d/o illuminating--geometrical viewing conditions. Diameter of the globe of diffusion is 150mm and diameter of the testing hole is 30mm. Light absorber is provide to eliminate the effect of mirror reflection.
10 successive test data auto-calculates mean value
With power-off protection function to protect data
The appearance of this instrument is novel and compact and the advanced circuit design effectively guarantees the accuracy and stability of the measurement data.
The instrument is equipped with a standard RS232 interface which can cooperate with the computer software for communication.
Adopt d/o illuminating--geometrical viewing conditions. Diameter of the globe of diffusion is 150mm and diameter of the testing hole is 30mm. Light absorber is provide to eliminate the effect of mirror reflection.
10 successive test data auto-calculates mean value
With power-off protection function to protect data
The appearance of this instrument is novel and compact and the advanced circuit design effectively guarantees the accuracy and stability of the measurement data.
The instrument is equipped with a standard RS232 interface which can cooperate with the computer software for communication.
National standard
GB/T 19281 、 GB/T 23774
Relevant information
NO